Saliva DNA Isolation Kit
Saliva DNA Isolation Kit
Register today to receive an exclusive 15% off* on your first order.
Supporting Data
Kit Specifications
|
|
Maximum Saliva Input
|
0.5 mL preserved saliva
0.25 mL fresh saliva |
Average Yield from 0.25 mL of Saliva*
|
3 - 7 μg
|
Average Purity (OD260/280)
|
1.7 - 2.1
|
Time to Complete 10 Purifications |
30 minutes
|
* Average yield will depending on the donor
Storage Conditions and Product Stability
All solutions should be kept tightly sealed and stored at room temperature. This kit is stable for 1 year after the date of shipment. The kit contains a ready-to-use Proteinase K, which is dissolved in a specially prepared storage buffer. The buffered Proteinase K is stable for up to 1 year after the date of shipment when stored at room temperature.
Component | Cat. RU45400 (50 preps) |
---|---|
Lysis Buffer F | 30 mL |
Proteinase K in Storage Buffer | 1.2 mL |
Binding Buffer B | 12 mL |
Wash Solution A | 18 mL |
Elution Buffer B | 15 mL |
Spin Columns | 50 |
Collection Tubes | 50 |
Elution Tubes (1.7 mL) | 50 |
Product Insert | 1 |
Documentation
The Range of DNA Yield with Norgen’s Saliva DNA Collection, Preservation and Isolation Kit
Determination of the DNA Molecular Weight (MW) from different Norgen Columns and Isolation Methods
Stability of DNA Stored in Norgen’s Saliva DNA Preservative for 52 Months at Room Temperature
Linearity of DNA Isolated from Increasing Volumes of Preserved Saliva Using Norgen’s Saliva DNA Isolation Kit
DNA Isolation from Saliva Preserved with Norgen’s Saliva DNA Collection and Preservation Device using Qiagen’s QIAamp DNA Blood
Comparative Study of DNA Isolated from Saliva Preserved in Norgen’s Preservative Using Norgen’s Saliva DNA Isolation Kit Versus
The Effect of Elution Volume on DNA Quantity and Quality Using Norgen’s Saliva DNA Isolation Kit
A Comparative Study Between Two Saliva DNA Preservation Systems, and Two Column-Based Saliva DNA Purification Methods
Column-Based Isolation of DNA from Preserved Saliva Samples Offers High Quality, Consistent, and Rapid DNA Isolation
Stability of DNA Stored in Norgen’s Saliva DNA Preservative at 55°C
Long Term Stability of DNA Stored in Norgen’s Saliva DNA Preservative
FAQs
Spin Column
Column clogging may occur due to one or combination of the following factors:
- Centrifugation speed was too low or spin time was inadequate.
Check the centrifuge to ensure that it is capable of generating the required RPMs. Sufficient centrifugal force is required to move the liquid phase through the column. Also ensure that the correct spin times are followed. Spinning for a few additional minutes will help. - The sample is too large.
Too many cells were applied to the column. Ensure that no more than 0.5 mL of preserved saliva is applied to the column. Clogging can be alleviated by centrifuging for a longer period of time until the lysate passes through the column. - The lysate/binding solution mixture is not homogeneous.
To ensure a homogeneous solution, vortex for 10- 15 seconds before applying the lysate to the spin column.
Low genomic DNA yield may result from one or combination of the following factors:
- Incomplete lysis of cells Increased.
Proteinase K incubation time at 55°C may result in increased yields. -
The DNA elution is incomplete.
Perform an additional centrifugation of 2 minutes at 14,000 x g to ensure that all the DNA is eluted. -
DNA concentration in the saliva sample being used is low.
Some saliva samples contain very little DNA. This varies from individual to individual based on numerous variables. Increased proteinase K incubation time at 55°C may result in increased yields.
Poor downstream performance of DNA may result from one or combination of the following factors:
-
DNA was not washed with Wash Solution A.
Traces of salt from the binding step may remain in the sample if the column is not washed with Wash Solution A. Salt may interfere with downstream applications, and thus must be washed from the column. -
Ethanol carryover.
Ensure that the dry spin after the column wash steps is performed, in order to remove traces of ethanol prior to elution. Ethanol is known to interfere with many downstream applications.
RNA might get co-eluted with DNA in some cases. Carry out a digestion with RNase A on the elution if the RNA present will interfere with downstream applications. Refer to manufacturer’s instructions regarding amount of enzyme to use, optimal incubation time and temperature.
Swab samples can be processed using Saliva DNA isolation kit following a supplementary proptocol. However, please note that swab samples are best processed using Norgen's Microbiome DNA isolation kit.
Yes, it has been tested on canine samples, and it is expected to work on most other non-human samples. Please contact our technical support team at support@norgenbiotek.com and ask for reference publications.
Yes, please contact our technical support team at support@norgenbiotek.com for help with supply of extra proteinase K in storage buffer.