Preserved Blood RNA Purification Kit II (for use with PAXgene Blood RNA Tubes)
For research use only and NOT intended for in vitro diagnostics.
Preserved Blood RNA Purification Kit II (for use with PAXgene Blood RNA Tubes)
Register today to receive an exclusive 15% off* on your first order.
Supporting Data
Kit Specifications
|
|
Maximum Column Binding Capacity
|
50 μg
|
Maximum Column Loading Volume |
650 μL
|
Size of RNA Purified |
All sizes, including
small RNA (< 200 nt) |
Time to Complete 10 Purifications |
30 minutes
|
Average Yield |
5 - 20 μg per 2.5 mL
preserved human blood |
Storage Conditions and Product Stability
All solutions should be kept tightly sealed and stored at room temperature. This kit is stable for 1 year after the date of shipment. The DNase I should be stored at -20°C.
Component | Cat. 43500 (50 preps) |
---|---|
NPX1 | 2 x 110 |
NPX2 | 40 mL |
NPX3 | 22 mL |
NPX4 | 6 mL |
NPX5 | 6 mL |
DNase I | 1 vial |
Mini Spin Columns | 50 |
Collection Tubes | 50 |
Elution Tubes (1.7 mL) | 50 |
Product Insert | 1 |
Documentation
FAQs
Spin Column
- Incomplete lysis of blood.
Ensure that the appropriate amount of NPX2 was used.
- An alternative elution solution was used.
It is recommended that the NPX5 Solution supplied with this kit be used for maximum RNA recovery.
- Ethanol was not added to the lysate.
Ensure that the appropriate amount of ethanol is added to the lysate before binding to the column.
- Ethanol was not added to NPX3.
Ensure that 50 mL of 96-100% ethanol is added to the supplied NXP3 prior to use.
- Insufficient centrifugation time and speed when pelleting the stabilized blood RNA before addition of NPX2 for Lysis.
Centrifuge the PAXgene™ Tube for 10 minutes at 3,000–5,000 x g.
- Low RNA Content in Sample.
Amount of cellular RNA could vary drastically in different individuals. In particular, blood from healthy individuals with low white blood cell count is expected to have lower RNA yield
- Insufficient solubilization of blood.
Ensure that the appropriate amount of NPX1 and NPX2 was used.
- Centrifuge temperature is too low.
Ensure that the centrifuge remains at room temperature throughout the procedure. Temperatures below 15°C may cause precipitates to form that can cause the columns to clog.
- RNase contamination.
RNases may be introduced during the use of the kit. Ensure proper procedures are followed when working with RNA. Please refer to “Working with RNA” at the beginning of the protocol's user guide.
- Procedure not performed quickly enough.
In order to maintain the integrity of the RNA, it is important that the procedure be performed quickly.
- Improper storage of the purified RNA.
For short term storage RNA samples may be stored at -20°C for a few days. It is recommended that samples be stored at -70°C for longer term storage.
- RNA was not washed 3 times with NPX3.
Traces of salt from the binding step may remain in the sample if the column is not washed 3 times with NXP3. Salt may interfere with downstream applications, and thus must be washed from the column.
- Ethanol carryover.
Ensure that the dry spin under the Column Wash procedure is performed, in order to remove traces of ethanol prior to elution. Ethanol is known to interfere with many downstream applications.